Radiation Output Accuracy

This test verifies that the stated kVp and HVL are within the margin of error of the detection equipment.

All measurements are taken with detector in contact with the face of the applicator.

Dose Linearity

This test was to ensure that dose linearly scaled with exposure time.

To do so, total dose measurements were taken with the E (50 mm dia., 50 mm SSD) and H (50 mm dia., 100 mm SSD) cones at 30 secs to 4+ mins exposure times at 0 mm depth.